In this experiment, you will explore the principles of acids, bases, and buffers by performing titration analysis.
Written by staff of ADInstruments.
Scientists quantify the “acidity” of an aqueous solution by expressing its molar concentration of hydronium ions (H3O+) on a logarithmic scale called the pH scale. The pH of an aqueous solution is calculated using the following equation:
Acidic solutions have a pH value of less than seven. Solutions with pH values greater than seven are described as basic, or alkaline. We use the pH of water as our definition of neutrality. Water is actually a mixture of molecular water (H2O), and ionized water (H3O+ and OH-). In pure water, the concentrations of H3O+ and OH- are in equilibrium at 10-7M. Therefore, pure water has a pH of 7.0. Solutions with high concentrations of hydronium ions have low pH values, while solutions with low hydronium ion concentrations have high pH values. It is important to note the intimate relationship of hydronium and hydroxyl ions. As one species becomes more prevalent, the other decreases in concentration. The pH values of several common substances are shown on the scale in Figure 1.
-), form an important biological buffer system. Bicarbonate ions can take up excess protons from the solution, while carbonic acid can donate protons to the solution. The bicarbonate buffer system is important in the blood.
In this experiment, you will determine the acid-base properties of solutions by performing a titration. Titration is a powerful analytical technique that can be used to determine concentration of unknown solutions. By adding a known volume of an acid or base to an unknown solution and measuring the change in pH, you can determine the molar concentration of that solution. This analysis is possible because hydronium and hydroxyl ions will neutralize each other. When the amount of hydronium ion in solution is exactly matched by the amount of hydroxyl, the pH of the solution will be 7.0. If you know how much acid or base is required to neutralize an unknown, then you can calculate the concentration of that unknown. You can also use a titration to examine the buffering properties of molecules by making a titration curve. A titration curve is a graph of the pH of a solution versus the quantity of titrant added. From a titration curve, you can determine the equilibrium constant, Ka, of an acid.
A computer system
PowerLab 4/25T
Chart, version 5.0.1 or later
pH Pod
pH Electrode
Reflective drop counter
Ring-stand
Ring clamp
Buret clamp
Buret with Teflon stopcock
250mL beakers
Magnetic stirring motor
Teflon stir bar
Electronic balance
Weighing boats
Graduated cylinders
100mL volumetric flask
10mL serological pipettes (disposable)
Wash bottle
Pipette bulb
Plastic funnel
Plastic drinking straw
pH standard solutions: pH 7.0, 4.0, and 10.0
Hydrochloric acid, 1.0N standard solution
Sodium hydroxide, 1.0N standard solution
L-histidine, hydrochloride salt
Sodium bicarbonate
Distilled water
Household vinegar
Safety
Warning: Always wear gloves and appropriate eye protection
when handling reagents.
1. Clamp the Drop Counter to your ring stand and position it so it is above the stir motor (Figure 2).
2. Place a 100mL beaker containing pH 7.0 buffer or distilled water on the stir motor.
3. Place the pH electrode into the holder on the Drop Counter so the tip is immersed in the buffer.
4. Connect the cable from the Drop Counter into the Input 1 Pod Port on the front of your PowerLab.
5. Connect the cable from the pH Pod into the Input 2 Pod Port on the front of the PowerLab.
6. Attach the BNC connector on the pH Electrode to the socket on the rear panel of the pH Pod.
7. Mount a 25 mL buret in the buret clamp on your ring stand (Figure 2).
8. Adjust the position of the buret tip so that it is aligned with the alignment marks on the Drop Counter (Figure 3). The buret tip should be positioned 3-5mm above the opening in the drop counter.
9. Close the stopcock on the buret.
Make sure your computer is turned on, and that the PowerLab is connected to it via its USB cable.
2. Launch Chart, and from the Experiments Gallery, open the file called “Titration Settings”.
3. After a few seconds, a blank Chart file will appear with two channels.
4. Fill your buret completely with distilled water.
5. Place a beaker under the drop counter and open the stopcock on the buret to allow water to fill the tip. Slowly run about 5mL of water through your buret to remove any air bubbles and then close the stopcock.
6. Remove the beaker and refill your buret. You are now ready to calibrate the Drop Counter.
7. Weigh a clean, dry 125mL Erlenmeyer flask and record the mass in Table 1 of your Data Notebook. Next, place the 125mL Erlenmeyer flask beneath the buret and drop counter. Leave the pH electrode in its beaker of buffer, and out of the way of the buret tip, for now.
8. In Chart, click Start.
9. Slowly open the stopcock so that a slow but steady stream of individual drops exit the buret. You should try to achieve a drop rate of one or two drops per second. Note: If the water exits the buret in a continuous stream, stop and repeat the calibration procedure.
10. Allow the buret to completely empty into your flask.
11. In Chart, click Stop.
12. Weigh the flask with the water. Record the weight in Table 1 of your Data Notebook.
13. Determine the weight of the water in the flask by subtracting the weight of the empty flask from the weight of the full flask. This weight is equal to the volume of water, in milliliters. Record your results in Table 1 of your Data Notebook.
14. In Chart, select the data in the Volume channel that has the endpoint of your calibration (Figure 4).
From the Volume channel function pop-up menu, choose Units Conversion. A dialog box will appear (Figure 5).
Obtain a beaker to use as a wastewater container.
2. Place 30-40mL of pH 4.01 buffer in a 100mL beaker.
3. Place 30-40mL of pH 10.01 buffer in another 100mL beaker.
4. Remove the pH electrode tip from its beaker, and rinse the tip with distilled water, using your “waste” beaker to catch the drips.
5. Place the pH electrode into the beaker of pH 4.01 buffer.
6. In Chart, click Start.
7. Record for 20 seconds.
8. Remove the pH electrode from the buffer, rinse the tip into your waste beaker with distilled water, and then replace the tip into the pH 10.01 buffer.
9. Record for 30 seconds.
10. Click Stop.
11. Rinse the pH electrode and return it to its beaker of pH 7.0 buffer.
12. Make a selection of your data in the pH channel.
13. From the pH channel function pop-up menu, choose, “pH”. A dialog box will appear (Figure 6).
14. Select the data in the left-hand window that corresponds to pH 4.01, and click the arrow next to Point 1.
15. Select the data corresponding to pH 10.01, and click the arrow next to Point 2.
Obtain a clean, dry 100mL volumetric flask.
2. Use a volumetric pipette to add 10mL of 1N NaOH standard to the flask.
3. Fill the flask to the fill line with distilled water.
4. Cap the volumetric flask, and invert it several times to mix the contents fully3.
In this exercise, you will titrate sodium hydroxide with hydrochloric acid to determine the equivalence point.
1. Make sure the stopcock on your buret is closed.
2. Fill the buret with 3mL of 0.1M NaOH standard that you prepared. Use the plastic funnel when filling your buret. Then partially drain the buret into a waste beaker marked “alkali waste” so that there are no bubbles in the buret tip. Close the stopcock when no air bubbles remain in the tip. Note: Label all your beakers so you do not inadvertently mix acids and bases.
3. Fill the buret to the top mark with 0.1N NaOH standard. Make sure there are no air bubbles in the buret tip; drain a little of the NaOH if you have to.
4. Obtain a clean, dry 250mL beaker. Place a Teflon stir bar into the beaker.
5. Add 50mL of distilled water to the beaker.
6. Pipette 1mL of 1N HCl standard into your beaker, and then place the beaker on the stir motor underneath the buret.
7. Rinse the pH electrode and place it into the beaker, making sure that the tip is immersed and not in contact with the stir bar.
8. Turn on the stirring motor, and set it to a slow speed.
9. In Chart, click Start.
10. Enter a comment called “Part 1” into your data trace.
11. Slowly open the stopcock on your buret until you get a drop rate of 1-2 drops per second.
12. Continue recording until the buret is completely empty.
13. When the buret is empty, click Stop.
14. Close the stopcock on your buret.
15. Save your data with an appropriate filename, such as “Acid-base titration”.
16. Follow the procedures in the Analysis section for determining the equivalence point of your titration.
Commercial vinegar contains dilute acetic acid. In this exercise, you will determine the concentration of acetic acid in vinegar by analytical titration.
1. Click the “New Chart Document” button in the Chart Toolbar Menu. If prompted, choose “Use the current settings” and make sure you have saved any previous data.
2. Pipette 5mL of household vinegar into a 100mL volumetric flask.
3. Fill the flask to the fill line with distilled water. This represents a 1:20 dilution of the vinegar.
4. Fill your buret with 0.1M NaOH standard.
5. Make sure there are no air bubbles in the buret tip; use your waste beaker if necessary to drain air bubbles from the tip.
6. Place a Teflon stir bar into a clean 250mL beaker, and pour 50mL of diluted vinegar into it.
7. Place the beaker under the buret; rinse the pH electrode and place it into the vinegar.
8. In Chart, click Start.
9. Enter a comment called “vinegar” into your data trace.
10. Slowly open the stopcock on your buret until you get a drop rate of 1-2 drops per second.
11. Continue recording until the buret is completely empty.
12. When the buret is empty, click Stop.
13. Close the buret stopcock.
14. Save your date with an appropriate filename, such as “Vinegar Titration”.
15. Follow the procedures in the Analysis section to determine the concentration of acetic acid in vinegar.
Bicarbonate ions help regulate the pH of biological fluids, especially blood. In this exercise, you will determine the pKa of bicarbonate, and then examine how bicarbonate helps regulate blood pH when CO2 is added to solution.
1. Click the “New Chart Document” button in the Chart Toolbar Menu. If prompted, choose “Use the current settings” and make sure you have saved any previous data.
2. Place 100mL of distilled water into a clean 250mL beaker and add a Teflon stir bar.
3. Weigh out 0.1g of sodium bicarbonate, and add it to the beaker. Make sure the solution is mixed well and the sodium bicarbonate is completely dissolved.
4. Pipette 1.3 mL of 1.0N HCl into your beaker.
5. Fill your buret with 0.1N NaOH standard, and make sure there is no air in the buret tip.
6. Make sure the stopcock on the buret is closed, and place the beaker on the stirring motor beneath the buret.
7. Rinse the pH electrode with distilled water and position it in your beaker.
8. Turn on the stirring motor to a slow speed.
9. Click Start to begin recording.
10. After ten seconds, slowly open the stopcock until you get a drop rate of 1-2 drops per second.
11. Continue recording until your buret is empty.
12. Close the buret stopcock and click Stop to end your data collection.
13. Remove the pH electrode from the beaker, rinse it, and return it to the beaker of pH 7 buffer.
1. Remove the buret from its holder.
2. Add 100mL of distilled water to a clean, dry 250mL beaker, and add a Teflon stir bar.
3. Place the beaker on the stir motor, rinse the pH electrode tip and place the electrode into the beaker of water.
4. Turn the stirring motor on and set it to a slow speed.
5. Click Start to begin recording. Add a comment to your trace called “DI water”.
6. Have one member of your group exhale into the beaker through a drinking straw. Try to exhale for 10-15 seconds. Enter a comment called “CO2” to your recording.
7. Click Stop, and observe your data.
8. After you observe your data, click Start again.
9. Add 0.1g of sodium bicarbonate to your water, and enter a comment called “bicarbonate”.
10. After the sodium bicarbonate is fully dissolved, perform the exhalation procedure by blowing into the beaker with a straw for 10 seconds. Enter a comment into your data trace called “CO2”.
11. Repeat the bubbling procedure three times, entering a comment each time.
12. Click Stop to end your recording.
13. Save the data with an appropriate filename, such as “Bicarbonate data”.
14. Refer to the analysis section for the analysis procedures.
In this exercise, you will make a complete titration curve of the amino acid histidine, and use your results to calculate its pKa values.
1. Click the “New Chart Document” button in the Chart Toolbar Menu. If prompted, choose “Use the current settings” and make sure you have saved any previous data.
2. Weigh out 0.1g of histidine HCl in a weigh boat using your electronic balance. Record the exact mass you used in your data notebook.
3. Add a Teflon stir bar to a clean, dry 250mL beaker.
4. Add the histidine HCl to your beaker.
5. Add 50mL of distilled water and mix the solution thoroughly.
6. Use a serological pipette to transfer 3.0mL of 1N HCl to the beaker.
7. Rinse your buret with 0.1M NaOH, and then fill the buret. Make sure there are no air bubbles in the buret tip. Make sure that you have at least 50mL of NaOH standard available for the titration.
8. Place the histidine solution underneath the drop counter; rinse the tip of the pH electrode with distilled water and place it into the beaker.
9. Turn on the stir motor to a slow speed. Do not proceed to the next step until you are sure the histidine is fully dissolved.
10. In Chart, click Start.
11. Enter a comment to your recording called “Histidine”.
12. Record for five seconds, and then slowly open the stopcock on your buret until you get a drop rate of 1-2 drops per second.
13. Continue recording until you reach a pH of at least 11.5. If you are using a 25mL buret, you may need to fill it with additional NaOH standard during your titration.
14. When the pH of your solution is 11.5, close the stopcock on your buret.
15. Save your data with an appropriate filename, such as “Histidine”.
16. Follow the procedures in the Analysis section to determine the pKa values of histidine.
1. Make a selection of your entire titration data set by clicking and dragging the mouse in the time axis at the bottom of the Chart view window.
2. Click the X-Y Plot button in the Chart toolbar.
3. In the X-Y window, click the number 1 box on the x-axis and the number 2 box on the y-axis (Figure). This will display pH versus volume.
5. Record the volume of NaOH added to reach the equivalence point.
6. Calculate the number of moles of NaOH added to neutralize the acid and then calculate the molar concentration of acid using the equations below.
Where V= volume, in mL
M= molarity, in moles per liter
1. Examine your recording from Exercise 2.
2. Using the mouse, select the entire data trace in both channels by clicking and dragging the time axis.
3. Click the X-Y Plot button in the Chart Toolbar.
4. In the X-Y window, click the number 1 box on the x-axis and the number 2 box on the y-axis. This will display pH versus volume.
5. Using the mouse, move the waveform cursor until you find the point on the curve where the pH is 7.0.
6. Record the volume of NaOH added to reach the equivalence point.
7. Calculate the number of moles of NaOH required to neutralize the acid, and then calculate the concentration of acetic acid (Hac) in vinegar using the equation below. Record your results in Table 3 of your Data Notebook.
1. Examine your data for Exercise 3.
2. Select the titration curve of sodium bicarbonate by clicking and dragging the mouse in the time axis.
3. Click the X-Y plot button in the Chart Toolbar.
4. In the X-Y window, select Channel 1 for the x-axis and Channel 1 for the y-axis.
5. Examine the titration curve. You may wish to use the “Print” function to print the X-Y window.
6. The titration curve should have one or more distinct “plateau” phases, where pH does not change rapidly. The pH value at the mid-point of each plateau corresponds to the pKa value. This is the pH at which the concentrations of conjugate base and undissociated acid are equal.
7. You can estimate the midpoint of the plateau phase as follows:
· In Chart, choose: SetupàChannel Settings. The Channel Settings dialog box will open (Figure 7).
· In the box that reads “number of channels”, enter “4”, and click OK to return to the Chart view.
· Two
new channels should now be visible in your Chart view. Click the Autoscale
button in the
Chart toolbar menu to scale the data in all channels. Channel 3 (pH rate)
shows the rate of change of the pH data in Channel 2 with the units of pH
sec-1. Channel 4 (Smoothed) is the same data as in Channel 3,
but with smoothing applied to reduce noise. You should observe two prominent
peaks in Channel 4 (Figure 8).
· Place the mouse cursor on the first large peak in Channel 4, and click the mouse button once to set the active point. Choose Commandà Add to Data Pad. Open the Data Pad by clicking the Data Pad button in the Chart toolbar. You should see two numbers entered: Volume of NaOH and pH. Enter the volume in Table 4b of your Data Notebook as V1.
· Repeat the procedure in the previous step for the second major peak in Channel 4. Record the volume NaOH in Table 4b as V2.
· Subtract V2 from V1 to get V3.
· Divide V3 by 2 to get V4.
· Add V4 to V1. This will be the midpoint estimate.
· Move the waveform cursor along your data in the NaOH channel until the number displayed is as close as possible to the midpoint estimate you calculated. Click the mouse button once to set the active point, and then choose Commandà Add to Data Pad. The pH value displayed in the data pad for this point is the estimated pKa value for bicarbonate.
Figure 7. Use the Channel Settings dialog box to add two new channels to your titration file.
Figure 8. The Zoom view for Channel 4 shows the change in pH over time (∂pH/∂t). The two large peaks in the differential curve can be used to indicate V1 and V2 for determining pKa.
8. Use the waveform cursor to complete Table 4b in your Data Notebook by recording the appropriate pH values from your breathing experiment.
1. Examine your data from Exercise 4.
2. Determine the pKa value(s) for histidine by following the procedures in Exercise 3, above, with the following changes:
· Determine V1, V2 and V3 in Channel 4, as you did in Exercise 3, above.
· Estimate pka1 as the pH at 1/2 V1.
· Determine the midpoint volumes for the curves between V1 and V2, and between V2 and V3. Estimate pKa2 and pKa3 in the same manner as you did for the bicarbonate buffer in Exercise 3.
3. Record the pKa value(s) in Table 5 of your Data Notebook.
Mass of empty flask (g)
Mass of full flask (g)
Mass of water in flask (g)
Volume of water (mL)
Number of drops
Drops per mL
Concentration NaOH (M)
Volume HCl (mL)
Volume NaOH added (mL)
Moles HCl in solution
Moles NaOH added
Concentration HCl (M)
Concentration NaOH (M)
Volume Hac (mL)
Volume NaOH added (mL)
Moles Hac
Moles NaOH added
Concentration Hac in beaker (M)
Concentration Hac in vinegar (M)
pKa1
pKa2
pKa3
1. A solution has a hydronium ion concentration of 1.8 x 10-3M. Calculate the pH of this solution.
2. A solution has a hydroxyl concentration of 3.4 x 10-8M. What is the solution’s hydronium ion concentration? What is the solution’s pH?
3. If pure water is left exposed to air, over time its pH decreases. What explanation can you give for this occurrence? Hint: Recall your results from Exercise 3.
4. What compound is formed when you exhale into water?
5. Compare the change in pH when you exhaled into water with when you exhaled into the bicarbonate solution. How can you explain your results?
6. Why do you suppose bicarbonate is a good buffer for blood?
7. How many pKa values did histidine have? At what pH value(s) could histidine be used as a buffer?
8. Draw the structure of histidine as it appears at:
pH 0.0 pH
4.0 pH 7.5 pH 10.5
Copyright © 2004 ADInstruments.
All rights reserved.
MacLab and PowerLab
are registered trademarks, and Chart and Scope are trademarks, of ADInstruments.
Windows and the Windows logo are either trademarks or registered trademarks
of Microsoft Corporation. Macintosh and the Mac logo are either trademarks
or registered trademarks of Apple Computer, Inc. Other trademarks are
the properties of their respective owners.
Contacts
International (Australia)
Tel: +61 (2) 9899
5455
Fax: +61 (2) 9899
5847
E-mail: info@adi.com.au
Web: www.adinstruments.com
North America
Tel: +1 (888) 965
6040
Fax: +1 (866) 965
9293
E-mail: info@adinstruments.com
Europe
Tel: +44 (1865) 891
623
Fax: +44 (1865) 890
800
E-mail: info@adi-europe.com
Japan
Tel: +81 (3) 5820
7556
Fax: +81 (3) 3861
7022
E-mail: info@adi-japan.co.jp
Asia
Tel: +86 (21) 5830
5639
Fax: +86 (21) 5830
5640
E-mail: info@adinstruments.com.cn